SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
5-2I2
5.3 SYSTEM SETTING PROCEDURES
5.3.1 GENERAL
This paragraph provides controller system setting procedure.
NOTICE: Please follow by the procedures of item 5.3.3. if it is unnecessary, please do not
change data of system setting to avoid hazards.
5.3.2 BRIEF INTRODUCTION
Controller system setting procedures listed below:
1. LED Lightness set
2. Air temperature sensor compensation
3. Air flow temperature sensor compensation
4. Skin temperature sensor compensation
5. Machine Number
6. Motor speed alarm value
7. Set value of Low deviation for air temperature
8. Set value of High deviation for air temperature
9. Set value of Low deviation for skin temperature
10. Set value of High deviation for skin temperature
11. Minimum set value of air temperature
12. Maximum set value of air temperature
13. Minimum set value of skin temperature
14. Maximum set value of skin temperature
15. Set value of Over temperature alarm (37 )
16. Set value of Over temperature alarm (37 )
17. Set value of air flow Over temperature alarm (37 )
18. Set value of air flow Over temperature alarm(37 )
7/31/2019 YP-90 Series Ajuste
2/22
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
5.3.3 PROCEDURE
At first, connected power supply, press and hold Set key for 2s as soon as switch on until Air
temprature indicator displays --.-, and no display on Skin and Set temprature indicator. And then,
press Up or Down until Air temprature indicator displays 78.6, after then, press Set key to enter into
the system set screen, or press Silence/Reset key to exit.
After entering into the set screen, Set temprature indicator displays set code, Air temprature
indicator displays the specific set data.
Turn on the power switch
5-3I2
Press and hold the for 3s
Enter password
Press key, EXIT Press or key to enter password 78.6,
Then press key to confirm
Error Password check
YES
System setting screen
Press key or Key, set temperature indicator displays
the set code
Press or Key can change
current set value
END
Press key Press and together
SAVE and EXIT Return to original set
Brightness set of indicator
Figure 5.2
7/31/2019 YP-90 Series Ajuste
3/22
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
5-4I2
CODE 00.1 LED Lightness set
Indicate the lightness , and set range is 00.1~01.5. The larger the lighter.
CODE 00.2Air temperature sensor compensation
Set value of temperature compensation for air temperature sensor , and set range is -5.0~05.0 ,
the first figure of set value means the increasing and decreasing value. For Example, 00.2 means
that the air temperature adds 0.2 as compensation. -1.3 means that the air temperature reduces
1.3 as compensation.
CODE 00.3Air flow temperature sensor compensation
temperature compensation for air flow temperature sensor , and set range is -5.0~05.0 , the
first figure of set value means the increasing and decreasing value. For Example, 00.2 means that
the air flow temperature adds 0.2 as compensation. -1.3 means that the air flow temperature
reduces 1.3 as compensation.
CODE 00.4 Skin temperature sensor compensation
Set value of temperature compensation for skin temperature sensor , and set range is
-5.0~05.0 , the first figure of set value means the increasing and decreasing value. For Example,
00.2 means that the skin temperature adds 0.2 as compensation. -1.3 means that the skin
temperature reduces 1.3 as compensation.
CODE 00.5 Machine Number
Set method of machine number, and set range is 00.0~02.0. For example, 00.1 measn No. 1
machine, 01.5 means No. 15.
CODE 00.6 Motor speed alarm valueSet method of fan speed, and set range is 00.0~25.5. Every 0.1 means 20 rpm. For example,
05.0 means 1000 rpm. Alarm will be active if the fan speed is lower than this value.
CODE 00.7 Set value of Low deviation for air temperature
Set method of low deviation alarm in air mode, and set range is 00.0~05.0. For example, 00.8
means 0.8 .
CODE 00.8 Set value of High deviation for air temperature
Set method of high deviation alarm in air mode, and set range is 00.0~05.0. For example, 00.8
means 0.8 .
CODE 00.9 Set value of low deviation for skin temperature
Set method of low deviation alarm in skin mode, and set range is 00.0~05.0. For example, 00.8
means 0.8 .
CODE 01.0 Set value of high deviation for skin temperature
Set method of high deviation alarm in skin mode, and set range is 00.0~05.0. For example, 00.8
means 0.8 .
7/31/2019 YP-90 Series Ajuste
4/22
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
5-5I2
CODE 01.1 Minimum set value of air temperature
Set method of Minimum set value of air temperature, and set range is 20.0~34.0. For example,
25.0 means Minimum set value 25.0 .
CODE 01. 2 Maximum set value of air temperature
Set method of Maximum set value of air temperature in mode of>37.0 , and set range is
37.0~39.0. For example, 39.0 means Maximum set value 39.0 .
CODE 01.3 Minimum set value of skin temperature
Set method of Minimum set value of skin temperature, and set range is 30.0~34.0. For example,
34.0 means Minimum set value 25.0 .
CODE 01.4 Maximum set value of skin temperature
Set method of Maximum set value of skin temperature in mode of>37.0 , and set range is
37.0~39.0. For example, 39.0 means Maximum set value 39.0 .
CODE 01.5 Set value of Over temperature alarm A (37
Set method of Over temperature alarm in air mode (set temperatuer37, and set range is
37.5~39.0. For example, 38.0 means over-temp alarm value 38.0 .
CODE 01.6 Set value of Over temperature alarm B (37)
Set method of Over temperature alarm in air mode (set temperatuer37, and set range is
38.5~40.0. For example, 40.0 means over-temp alarm value 40.0 .
CODE 01.7 Set value of air flow Over temperature alarm A (37
Set method of alarm value for air flow blocked (set temperatuer 37, and set range is
38.5~48.0. For example, 45.0 means alarm value for air flow blocked 45.0
CODE 01.8 Set value of air flow Over temperature alarm B (37)
Set method of alarm value for air flow blocked (set temperatuer >37, and set range is 40.5~52.0.
For example, 45.0 means alarm value for air flow blocked 45.0
7/31/2019 YP-90 Series Ajuste
5/22
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
5.4 TROUBLESHOOTING PROCEDURES
5.4.1 GENERAL
Troubleshooting guides for the Controller are provided in Paragraph 5.4.3 and Paragraph 5.4.4.
Paragraph 5.4.3 provides the alarm code when the controller alarms. Paragraph 5.4.4 provides
troubleshooting in the form of flowcharts.
It is confirmed that the system settings are correct before troubleshooting.
5.4.2 TEST EQUIPMENT REQUIRED
The test equipment listed below is required for troubleshooting the Controller. Equivalent test
equipment may be substituted.
Digital VOM, FLUKE 8060A
31.0 Air Temperature Simulation
5-6I2
Resistor
Resistor R=23166/0.25W/0.5%
31.0 Air flow Temperature Simulation
Resistor
Resistor R=23166/0.25W/0.5%
31.0 skin Temperature Simulation
Resistor R=23166/0.25W/0.5%
5.4.3 TROUBLESHOOTING
All alarms have the relevant alarm code except for the power ailure alarm, please refer to Table 5.1.
TABLE 5.1 ERROR CODES
ALARM CODE TYPEEXPLANATION AND
CORRECTIVE ACTION
E0.1 Air temperature sensor failure alarm Refer to Flow chart 5.1
E0.2 Isolated temperature sensor alarm Refer to Flow chart 5.2
E0.3 Air temperature deviation failure alarm Refer to Flow chart 5.3
Resistor
7/31/2019 YP-90 Series Ajuste
6/22
SERVICE MANUAL FOR I FANT INCUBATOR EDITION/REVISION B/0N
5-7I2
TABLE 5.1 ERROR CODES (Continued)
ALARM CODE TYPEEXPLANATION AND
CORRECTIVE ACTION
E0.4 Air flow sensor alarm Refer to Flow chart 5.4
E0.5 Skin temperature sensor failure alarm Refer to Flow chart 5.5
E0.6 Over temperature failure alarm Refer to Flow chart 5.6
E0.7 Air circulation failure alarm Refer to Flow chart 5.7
E0.8 Wrong position of skin sensor alarm Check the position of skin sensor
E0.9 Fan motor stop alarm Refer to Flow chart 5.8
E1.0 High deviation temperature failure alarmCheck whether there is heat sourcenearby, if yes, please leave far awayfrom the heat source.
E1.1 Low deviation temperature failure alarmCheck the close condition of all doors
and panels
H0.1 ROM failure alarm inside of Main MCU Replace Main MCU
H0.2Internal system failure alarm inside ofMain MCU
Replace Main MCU
H0.3 communication failure alarm
1. Check the connection of circuitbetween the main MUC and auxiliaryMCU2. Replace main MUC and auxiliaryMCU
H0.4 ROM failure alarm inside auxiliary MCU Replace auxiliary MCU
H0.5
Internal system failure alarm inside
auxiliary MCU Replace auxiliary MCU
H0.6 E2PROM failure alarm Replace Main MCU
H0.7 SRAM failure alarm Replace U13 on mian board
H0.8 Main ADC failure alarm Refer to Flow chart 5.9
H0.9 Auxiliary ADC failure alarm Refer to Flow chart 5.10
H1.0 Heat system failure alarm Refer to Flow chart 5.11
H1.1 Real timer RTC failure alarm Replace U20 on mian board
H1.2 Keyboard fault alarm When the key board of the control panelon controller short-circuit alarms, pleasereplace the control panel.
H1.3 Battery failure alarm
Inner storage battery failure alarm:1. Check the wire connection betweenthe inner storage battery and the powerboard.2. Replace the inner storage battery.
Power failure alarm Refer to Flow chart 5.12
7/31/2019 YP-90 Series Ajuste
7/22
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
5.4.4 TROUBLESHOOTING FLOW CHART
Before troubleshooting, forst, please switch off, and disconnect the power cable, and then open the
top cover of warmer module and the back cover of column. And then connect the power cable and turn
onthe switch, at the same time, make sure that power +5V, +12V of controller, element, and all cables
are in good condition.In working state, press Up Key and Air mode Key , you can find the isolated
temperature and air flow temperarture in the air and set temperature indicator respectively. So it is easy
for service man to find the cause of trouble.
AIR TEMPERATURE SENSOR FAILURE ALARM E0.1
no
Good conection? Check connection of air temperature sensor,
and the wiring of all circuit board
The 31.00.2 Air Temperature Simulation (decade
Equal to 23166182)is instead of the air
temperature sensor, turn on power supply
the temperature display is
31.0 and add the no
compensation temperature alarm E0.1? Replace the air temperature sensor
of system set procedure
yes
on the main board U11-6 is 2.3V? no check the connect to R21, U1-6
Yes
CHECK U11
FLOW CHART 5.1 AIR TEMPERATURE SENSOR FAILURE
5-8I2
7/31/2019 YP-90 Series Ajuste
8/22
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
ISOLATED TEMPERATURE SENSOR FAILURE ALARM E0.2
``
no
Good conection? Check connection of isolated temperature
sensor, and the wiring of all circuit board
The 31.00.2 Air Temperature Simulation (decade
Equal to 23166
182
)is instead of the isolatedtemperature sensor, turn on power supply
See 5.4.4, the isolated no
temperature is 31.0 alarm E0.2? Replace the isolated temperature sensor
yes
no
on the main board U21-6 is 2.3V? check the connect to R16, U21-6
Yes
Check U21
FLOW CHART 5.2 ISOLATED TEMPERATURE SENSOR FAILURE
5-9I2
7/31/2019 YP-90 Series Ajuste
9/22
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
Air temperature deviation failure alarm E0.3
``
noGood conection? Check connection of isolated temperature
sensor, and the wiring of all circuit board
The 31.00.2 Air Temperature Simulation (decade
Equal to 23166182)is instead of the air and isolated
temperature sensor, turn on power supply
See 5.4.4, check whether
The air temeprature and no
isolated tempertire is alarm E0.3? Replace sensor
added the temperature
compensation based on 31.0
yes
U11-6, no
on the main board U21-6 is 2.3V? check the connect to R16 and U21-6,
R21 and U11-6
Yes
Replace U11 or U21
FLOW CHART 5.3 AIR TEMPERATURE DEVIATION FAILURE
5-10I2
7/31/2019 YP-90 Series Ajuste
10/22
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
Air flow sensor failure alarm E0.4
noGood conection? Check connection of sensor,and the wiring of all
circuit board
The 31.00.2 Air Temperature Simulation (decade
Equal to 23166182)is instead of air flow
temperature sensor, turn on power supply
See 5.4.4, check whether
The air temeprature and no
isolated tempertire is alarm E0.4? Replace air flow sensor
added the temperature
compensation based on 31.0
yes
no
on the main board U11-8 is 2.3V? check the connect to R23 and U11-8
Replace U11
FLOW CHART 5.4 AIR FLOW SENSOR FAILURE
5-11I2
7/31/2019 YP-90 Series Ajuste
11/22
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
SKIN TEMPERATURE SENSOR FAILURE ALARM E0.5
Check connection between skin sensor
Good conection? Socket and circuit board, the connection Of all
circuit board
The 31.00.2 Air Temperature Simulation (decade
Equal to 23166182)is instead of air flow
temperature sensor, turn on power supply
the temperature display no
is 31.0, added alarm E0.5? Replace the skin temperature sensor
the temperature compensation
in system set procedure
yes
no
on the main board U11-7 is 2.3V? check connect Relay R22 to U1-7
Yes
Check U11
FLOW CHART 5.5 SKIN TEMPERATURE SENSOR FAILURE
5-12I2
7/31/2019 YP-90 Series Ajuste
12/22
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
Over-temp failure E0.6
yes
Is there heat source nearby? Far away from the heat source
5-13I2
no
yes
Air and isolated temperature Check air and isolated temperature sensor
Sensor is OK?
no
Replace air and isolated temperature sensor
FLOW CHART 5.6 OVER TEMPERATURE FAILURE
Check whether the inlet
Or outlet is blocked
Air circulation failure alarm E0.7
Check the air flow condition of inlet and outlet
FLOW CHART 5.7 AIR CIRCULAATION FAILURE
7/31/2019 YP-90 Series Ajuste
13/22
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
Fan Motor Failure Alarm E0.9
Motor rotating no Motor input terminal no Check the power cord
Is AC power?
yes
yes
on the power Pulse at no Check the Capacitor no
supply board U5-2? fan detected CM is good Replace the capacitor CM
board
yes yes
5-14I2
no Replace the Motor
on the power Pulse at Check U5
supply board U5-3
yes
on the main no
board Pulse U12-16? check the connection between the main board
and the power supply board
yes
Check U12
FLOW CHART 5.8 FAN MOTOR FAILURE ALARM( SHEET 1 OF 2)
7/31/2019 YP-90 Series Ajuste
14/22
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
Slow Motor Speed Alarm E0.9
yes
Motor input no
Voltage is ok? Check the power input
yes
no
Capacitor Replace the capacitor CM
CM is good
yes
Replace the motor
FLOW CHART 5.8 FAN MOTOR FAILURE ALARM ( SHEET 2 OF 2)
5-15I2
7/31/2019 YP-90 Series Ajuste
15/22
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
Connection of U11
and U12 is OK?
Check the jumper wire of
U11-13, R5, R6
noPin of U11-13 is
+2.5V?
no
Check R5, R6Crosspoint of R5
and R6 is +2.5V?
Main ADC failure alarm H0.8
On main board
On main board
Check the connecting wireOn main board
Replace U11
FLOW CHART 5.9 MAIN ADC FAILURE ALARM
5-16I2
7/31/2019 YP-90 Series Ajuste
16/22
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
5-17I2
Check the connecting wireConnection of U21and U18 is OK?
Check the jumper wire of
U21-9, R14, R15
noPin of U21-9 is
+2.5V?
no
Check R14, R15Crosspoint of R14
and R15 is +2.5V?
Auxiliary ADC failure alarm H0.9
On main board
On main board
On main board
Replace U21
FLOW CHART 5.10 AUXILIARY ADC FAILURE ALARM
7/31/2019 YP-90 Series Ajuste
17/22
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
Heat circuit failure alarm H1.0
Disconnet power supply, measure resistance of twoterminals for heater
5-18I2
yes
Replace heater
no
Heater is open circuit or
short circuit?
Conneat power board and main board
Element number from here is on
power board
Connect pin 1 (ssr-ctr) and pin 3
(relay-ctr) on J8 to VCC power make two
conrol signals high level
no
Base of triode Q2, Q3
is >0.7V?
Check triode Q2, Q3
and resistor R8,R9
yes
nothere +12V on solid
relay and control
terminal of relay K1
Replace triode
Q2 or Q3
yes
To sheet 2
FLOW CHART 5.11 HEAT SYSTEM FAILURE (SHEET 1 OF 4)
7/31/2019 YP-90 Series Ajuste
18/22
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
From sheet 1
Output terminal of solid
relay and relay K1 can
work?
no Replace solid relay
or relay K1
yes
Voltage of U1-2 is
> 5mV
yes
no
Check inductor T1
Base of triode Q1
>0V?
yes
noBase of triode Q1
>0.7V?
yes
no
Voltage of U1-1
is >2V?Check U1,R1 ,C4
Check R2,R6,C5
no
Repalce triode Q1
Connet pin 1 (ssr-ctr) and pin3 (relay-ctr) on J8 to GND
make two control signals high level
yes
To sheet 3
FLOW CHART 5.11 HEAT SYSTEM FAILURE (SHEET 2 OF 4)
5-19I2
7/31/2019 YP-90 Series Ajuste
19/22
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
From sheet 2
Base of triode Q1,Q3
is 0V?
yes
Check triode Q2,Q3
and resistor R8, R9
no
Base of triode Q1
is 5V?
yes
noBase of triode Q1
is 0V?
Yes
no
Voltage U1-1 is
0V?
Yes
no
Voltage U1-2 is
0V?
no
yesIs there +12V on
control terminal of solid
relay and relay K1
Replace triode
Q2 or Q3
Check inductor T1
Check U1,R1,C4
Check R2, R6, C5
no
Replace triode Q1
yes
To sheet 4
FLOW CHART 5.11 HEAT SYSTEM FAILURE (SHEET 3 OF 4)
5-20I2
7/31/2019 YP-90 Series Ajuste
20/22
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
From sheet 3
The pins on porer
board J8-1, 3, 4 and
J2-1, 3, 4 on main
board works?
Switch off during measuring
no
Replace connecting wire
yes
Connect power board and main board
Connect pins U8-2, 11 to VCC power
Element number from here is onmainboard
Yes
Pin of U8-9, 18 is
5V?
no
Check U8
Measure the voltage of U15-2 for self-checking
yes
no
Replace U15
yes yes
Replace main board
Still alarming after
replacing U15
Voltage is changingfrom low level to highlevel?
FLOW CHART 5.11 HEAT SYSTEM FAILURE (SHEET 4 OF 4)
5-21I2
7/31/2019 YP-90 Series Ajuste
21/22
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
Power Failure Alarm
5-22I2
Is there AC power on
input socket?Check outside power input
No
Yse
Initial trans for mer is
AC ower?
No Check the connecting wire of
trans for mer and ower cord,
and Fuse F1, F2, F3 on AC
entry board
Yes
No
Pin 3 on U2, U3 is +5V?
No
Cross point of R17,R18 is +2.5V?
No
Check U2, U3 and prefix circuit
Yes
On power board
Pin4 of U4 is
+12V?Check U4 and prefix circuitOn power board
Yes
Check R17, R18
Yes
Secondary output oftransformer is 9V, 15V? Replace transformer
On power board
Yes
To sheet 2
FLOW CHART 5.12 POWER FAILURE ALARM (SHEET 1 OF 2)
7/31/2019 YP-90 Series Ajuste
22/22
SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
From sheet 1
NoCrosspoint of
R19, R20 is 2.5V?On power board Replace R19, R20
Yes
Crosspoint of
R21, R22 is
2.1V?
No
Replace R21, R22On power board
5-23I2
Yes
Pins7, 1 of U6
is high level?
No
On power board Replace U6
Yes
NoPin5 of U8 is low
level?Check the connection of power board
and main board
On main board
Yes
Pin4 of U8 is
low level?
No
Replace U8On main board
Yes
Replace main board
FLOW CHART 5.12 POWER FAILURE ALARM (SHEET 2 OF 2)